Products
$50.00Bioview (Box of 25)
25 Tests with Crusher and Record Sheet
In-Office Biological Spore Testing
Bioview is a self-contained Biological Indicator for use in Steam sterilization cycles. Each unit is made up of a plastic container with cap, a crushable glass ampoule with recovery media, and a paper strip inoculated with spores. Bioview is available for use in dental and medical offices.
24-Hour Incubation
Intended for use in steam 121°C gravity and 121-135°C prevacuum cycles. Suitable for use in a dental or medical office. Contains G. stearothermophilus spores.
The control unit should exhibit turbidity and/or a color change to or toward yellow. If the control unit does not show signs of growth, consider the test invalid.
A failed sterilization cycle is indicated by turbidity and/or a color change to or toward yellow. A test unit that retains its original color, i.e., purple for Bioview Steam indicates that sterilization parameters have been met.
Each lot of Bioview is certified for population, species, D-value, z-value (where applicable), purity and expiry.
Compare Bioview® to 3M Attest™. Bioview is the affordable option. (Bioview® indicators can be used in Attest™ incubators.)
For orders outside the U.S. call 231-922-2211
or email sales@nbbs.com